最新文章专题视频专题问答1问答10问答100问答1000问答2000关键字专题1关键字专题50关键字专题500关键字专题1500TAG最新视频文章推荐1 推荐3 推荐5 推荐7 推荐9 推荐11 推荐13 推荐15 推荐17 推荐19 推荐21 推荐23 推荐25 推荐27 推荐29 推荐31 推荐33 推荐35 推荐37视频文章20视频文章30视频文章40视频文章50视频文章60 视频文章70视频文章80视频文章90视频文章100视频文章120视频文章140 视频2关键字专题关键字专题tag2tag3文章专题文章专题2文章索引1文章索引2文章索引3文章索引4文章索引5123456789101112131415文章专题3
当前位置: 首页 - 正文

基于DDS的精密频率测量的一种替代方法

来源:动视网 责编:小OO 时间:2025-10-01 09:46:46
文档

基于DDS的精密频率测量的一种替代方法

AnalternativemethodofprecisefrequencybytheaidofaDDSContentsAmethodoffrequencymeasurementbasedonaclosedloopcomposedmainlyofaFrequencyComparator(FC)andaDirectDigitalSynthesizer(DDS)ispresentedinthispaper.TheDDSservesasreferencesinewavesignalgeneratora
推荐度:
导读AnalternativemethodofprecisefrequencybytheaidofaDDSContentsAmethodoffrequencymeasurementbasedonaclosedloopcomposedmainlyofaFrequencyComparator(FC)andaDirectDigitalSynthesizer(DDS)ispresentedinthispaper.TheDDSservesasreferencesinewavesignalgeneratora
An alternative method of precise frequency by the aid of a DDS Contents

 A method of frequency measurement based on a closed loop composed mainly of a Frequency Comparator (FC) and a Direct Digital Synthesizer (DDS) is presented in this paper. The DDS serves as reference sinewave signal generator acting at one of the FC's inputs. The FC accepts the hard-limited waveform of the DDS as well as the unknown frequency. From the comparison of the two signals a logic output that controls an up/down counter is produced. The counter's output acting as the Frequency Setting Word (FSW) instructs the DDS to produce a new sinewave closer in frequency to the unknown one. When the loop settles, the FSW gives the digital estimate of the unknown frequency. Advantage is taken from the inherent high resolution of the DDS and noise immunity of the loop, to design an equally precise and immune frequency meter. All the additional associated stages up to the instrument's display are presented.

 1 Introduction

 The most commonly used frequency measurement technique adopts counters that count the pulses of the unknown frequency during a predefined time window (aperture). Apart from this, techniques where the pulses of a reference frequency are counted during one or more periods of the unknown one are also common. In the latter case, the period instead of the frequency is estimated .Some papers in [1] in the literature deal with the problem of low frequency measurement and are focusing in the frequency range of cardiac (heart) signals (a few hertz) or in the mains frequency (50-60 Hz).These techniques are actually measuring the period of the signals and use some way to calculate its reciprocal, the frequency. In [2], the frequency is calculated by the method of look-up tables. Others [4-6] are microprocessor or microcontroller based. 

The above methods can be characterized as open-loop methods i.e. digital counters are used to count during a predefined tinle interval and calculate the result afterwards. Its closed-loop form characterizes the proposed method in this paper. By the term "closed-loop" we denote some sort of feedback. A waveform with a known (controlled) frequency is produced within the circuit and is fed back to the frequency comparison stage which consecutively forces it to approximate the unknown (input) frequency. The device that produces the above mentioned waveform of controlled frequency is a Direct Digital Synthesizer.

 2 Direct Digital Synthesis

 A typical Direct Digital Synthesizer consists of a RAM containing samples of a sinewave (sine look-up table, LUT). These samples are swept in a controlled manner by the aid of a Frequency Setting Word (FSW), which determines the phase step. A typical FSW is 32-bit wide, but 48-bit synthesizers leading in higher frequency resolution are also available. A phase accumulator produces the successive addresses of the sine look-up table and generates a digitized sine wave output. The digital part of the DDS, the phase accumulator and the LUT, is called Numerically Controlled Oscillator (NCO). The final stage, which in contrast to the previous one is mostly analog, consists of a D/A converter followed by a filter. The filter smoothes the digitized sinewave, producing a continuous output signal. In the applications where a square wave output is needed, this is obtained by a hard limiter after the filter. It is not equivalent to use e.g. the MSB of the accumulator's output instead of the filtered and hard limited waveform because significant jitter will be encountered.

 The frequency of the output signal for an n-bit system is calculated in the following way; If n the phase step is equal to one, the accumulator will count by ones, taking 2 clock cycles to address the entire LUT and to generate one cycle of the output sinewave. This is the lowest frequency that the system can generate and is also its frequency resolution. Setting the FSW equal to two, results in the accumulator counting by twos, taking clock cycles to complete cycles of the output sinewave. It can easily be shown that for any integer m, where   , the number of clock cycles taken to generate one cycle of the output sine wave is  , and the output frequency (fDDS) and the frequency resolution (fres) are given by the following formulas:

                                          

                                           

For n = 32 and having a clock frequency of fclk = 33 MHz, the frequency resolution is 7.68 mHz. If n is increased to 48, with the same clock frequency, a resolution of 120 nHz is possible.

 3 The proposed frequency measurement technique

 The idea that led to our present design came from the extremely high frequency resolution of the DDS devices and is enforced by the noise immunity of its closed loop form. A (known) frequency source, the DDS, is employed in a closed loop and is forced progressively to produce an output with a frequency equal to the unknown input . A rule of thumb in the DDS systems is that the maximum acceptable synthesized frequency is about 25% of the clock frequency (well below the Nyquist limit). According to this, our prototype that uses a 33 MHz clock would effectively count up to 8 MHz. Looking at the GaAs products, we can see that recently available DDS devises can operate at clock frequencies up to the extent of 400 MHz. Therefore, by the present method, frequency counters working up to 100 MHz can be designed. The resolution will depend on the number of FSW bits and the clock frequency. The clock frequency fclk of the DDS is very critical because as it decreases, the resolution of the proposed method (defined as fclk/ 2n ) becomes finer i.e. it improves. The impact of the clock frequency decrease is the subsequent decrease of its maximum output frequency that limits the counter's maximum count. The major blocks have been shown . Among them are the Frequency Comparator and the DDS. To overcome some disadvantages of the specific frequency comparator a correction stage has been incorporated. This stage is also used for the measurement extraction in order to display the correct reading.

 3.1 Operation of the circuit 

The circuit operates in such a way that at the beginning of a new measurement the DDS output frequency would be controlled in a successive approximation way. The initial DDS frequency would be half of it's maximum. In addition, the frequency step of the approximation would equal the 1/4 of the DDS maximum frequency. On every approximation the frequency step is divided by two and added or subtracted to the FSW of the DDS, depending on the output of the Frequency Comparator. The approximation procedure stops when the step size decreases to one. After that, an up/down counter substitutes the approximation mechanism. 

The digital FSW, after the appropriate correction and decoding, is presented in an output device i.e. an LCD display or any other suitable means. Alternatively, it can be digitally recorded or it can be read by a computer.

 As conclusion of this initial approach we could say that the proposed method is based on a Digital Controlled Synthesizer which is forced to produce a frequency almost equal to the unknown one. 

3.2 Frequency comparison

 The frequency comparator seems to be the most critical stage of the design. The implementation is based on a modified phase/frequency comparator proposed by Philips in the 74HC4046 PLL device. It consists primarily of two binary counters, counting up to two and an RS flip-flop. The function of the frequency comparator is based on the principle that the lower frequency, i.e. larger period, includes (embraces) at least one or more full periods of the higher frequency (smaller period). This means that two or more rising edges of the higher frequency waveform are included within the lower frequency period. Considering the above, the circuit operates as follows: When the first counter (#1) encounters two rising edges of the unknown frequency in one period of the DDS, it sets the output of the RS flip-flop. The logic "1" of the RSflip-flop acting at the U/D control input of the Up/Down counter forces the DDS to rise its output frequency. On the contrary, when the second counter (#2) counts two rising edges of the DDS output within a period of the unknown frequency it resets the RS flip-flop's output. This action decreases the frequency of the DDS.

 At a first glance one could think that the synthesized frequency could reach the measured one (fin) and then the operation of the counter stops. Unfortunately this is not the case. A dynamic mechanism takes place instead. The circuit needs some time to realize the correct frequency relation. We will refer to this time as "hysteresis". Hysteresis depends on the initial timing relation of the DDS output and on the unknown frequency. Initially, during the hysteresis period, the indication regarding the larger frequency is ambiguous i.e. it can be erroneous. The ambiguity settles when two rising edges of the higher frequency waveform occur during one period of the lower one. If we consider the case of the DDS frequency to be equal to the unknown one, we will find that the comparator's output will toggle, indicating alternatively that the DDS frequency is higher or lower than the unknown. This is actually an acceptable and expected condition, because (as in a voltage comparator) an equality indication could not exist. In our case this is not a problem because the circuit is embedded in a closed loop. The loop will act in a manner that after some short time, the hysteresis, the situation will be reversed and so on. The duration of hysteresis is variable. This situation is controlled, as will be explained later. Although an analog implementation of the frequency comparator would look more robust to noise we insisted to the digital implementation for three reasons: ease of implementation in VLSI or Programmable Logic Devices (PLDs) with no need of analog components, wide frequency range of operation and shorter response time. 

3.3 Interaction between frequency comparator and digital synthesizer

 After the successive approximation of the unknown frequency the Frequency Comparator "realizes" that the synthesized frequency is higher (lower) than the unknown one and produces a logic 0 (1) at the output which commands the up/down counter to count in the down (up) direction. As previously mentioned, the output of this counter is considered to be the FSW to the DDS stage. In the case when the DDS frequency was initially lower, the synthesized frequency will increase progressively to reach the unknown one. This will not be "realized" by the frequency comparator and the synthesized frequency will keep on increasing for some clock cycles, until the comparator detects the correct relation of it's two input frequencies, the unknown one and the DDS output. The same phenomenon will be observed for the opposite (decreasing) case also. This is due to hysteresis that was mentioned earlier.

 When DDS output (fDDS) has approached fin, due to hysteresis, no specific frequency is synthesized. Instead, it swings between f1 and f2, where f1 and f2 are the two extreme values of the frequency swing lying symmetrically around fin. The DDS output can be considered as a frequency modulated carrier by a triangular waveform. The triangular waveform is the analog representation of the FSW applied to the DDS. lower trace shows a typical output of the Frequency Comparator. In the same figure, upper trace, is shown in analog form the FSW variation as it is trying to approach the correct value. This waveform has been captured using an auxiliary hardware circuit: A digital-to-analog converter (DAC) was connected to the output of the U/D counter (MSBs) in order to study the operation. This DAC is not shown in the block diagram of the circuit. Stated differently, the lower trace is the U/D command (input) to the counter while the upper trace is a hypothetical "frequency modulating" waveform. It is obvious that the term "hypothetical" is used because there is not such a waveform available somewhere in the circuit (except for the auxiliary DAC). Instead, its numerical equivalent exists. The magnitude of the slope of the elements of the triangular waveform is constant for constant input frequency and depends on the clock of the U/D counter (horizontal axis) and the voltage reference of the DAC (vertical axis). This slope is ±.

 3.4 Description of the prototype hardware

 For evaluation purposes two prototypes have been built and tested in the laboratory. The first approach was a low frequency instrument (operating up to 15 KHz) . The purpose of this implementation was to study the principles of operation of the proposed method. Next, a higher frequency prototype was built which will be described in more detail here. In order to implement the digital part of the prototype, (Frequency Comparator, Successive Counter, Correction Stage) two PLD devices from Altera (EPF 80LC68-12) were used. These devices are interconnected with the DDS, which is the Q2240I-3S1 from Qualcomm. The DDS has a 32-bit input and a 12-bit output for the sine lookup table (LUT). The 12-bit output of the LUT is fed into the D/A converter, the AD9713B from Analog Devices. Its analog output is connected to an I/V amplifier (current-to-voltage converter). 

The generated sinewave has upper harmonics, due to the DAC operation. These harmonics are removed from the filters that follow the DAC. The correction stage is implemented partially on the PLDs and partially on the microcontroller. Based on the up-down command of the frequency comparator we store the two extreme values, FSW1 and FSW2, which are then transferred into the micro-controller (Atmel ATC52), transformed into numerical representation and fed to the LCD Display. The micro-controller also controls the whole operation of the prototype.

 The behaviour of the instrument was according to the expected and was alike to a conventional bench frequency counter. The speed of measurement was checked using lower trace, obtained by the aid of a digital oscilloscope. Each state, high or low, of this waveform corresponds to the time required for one measurement. 

4 Conclusion 

In this paper an alternative method of frequency measurement has been proposed. It has been pointed out that in most cases this method is faster than conventional methods for the same frequency resolution. On the other hand, the precision of the method can be very high due to the inherent high frequency resolution characteristic of the DDS that is employed. This synthesizer, which can be thought as an oscillator, is driven to "oscillate" in the region of the unknown input frequency. A comparison with conventional methods has been given and two prototypes have been built and tested in the laboratory.

 The second major advantage of this method is that if repetitive frequency measurements are to be taken, the instrument remains locked and the frequency measurement does not restart from the beginning, but instead is automatically driven to lower or higher values. In other words, the loop has the capability to follow the changes in the frequency of the input signal. In the conventional counting techniques the counting procedure is repeated (restarted) for each new measurement. 

 Another important advantage is the noise immunity of the system, due to its closed loop nature. A detailed study of the noise behavior has not been carried out in this paper. This is mainly because the aim of this text is to present an alternative principle of frequency measurement. Moreover, the final output of the system is taken after some further processing (measurement correction) which also contributes to the noise immunity. 

借助 DDS 的精密测量频率的一种替代方法

 内容  

本文所提出的频率测量的方法基于闭环组成,主要是由一个频率比较器(FC)和直接数字频率合成 器(DDS)组成,对此在本文中进行了介绍。DDS 作为标准正弦波信号发生器在 FC 的输入之中扮演重要的角色。FC 接受了 DDS 的硬限幅波形以及未知的频率。从而依靠比较两 个信号的逻辑输出,控制向上/向下计数器的产生。计数器的输出作为频率设定字 (FSW) 指示DDS 产生一个新的正弦波,这个正弦波的频率接近未知频率之一。 当循环稳定, 频率设定字给出了未知频率的数字估计。从DDS所固有的具有高分辨率和环路噪声免疫的优势方面,设计同样精确和不受影响的频率计。所有额外相关的阶段都被仪器的显示器显示出来。

 1 简介

 最常用的测频技术是采用计数器在预定的时间窗口(光圈)测量未知频率的脉冲。此外,在一个或多个未知的时期内测量参考脉冲的技术也很常见。在后一种情况下,代替频率的周期只是估计的。本文献的第[1]部分的某些文件集中处理了在中心信号的频率范围(几赫兹)或电源频率范围(50-60 赫兹)之内的低频率的测量问题。这些技术实际上是在测量信号的周期,并使用一些方法 来计算它的倒数,即频率。在第[2]部分中,频率由查找表的方法计算。其他[4-6]的内容是以微处理器或微控制器为基础的。 

上述方法的特点是实用开环方法,即数字计数器,用来计算在预定 tinle 间隔之后的 结果。其闭环形式刻画了本文提出的方法。这个术语“闭环”我们用来记一些反馈。一个已知(控制)的频率波形在电路中产生,并反馈到比较阶段强制它来接近未知 的(输入)信号的频率。设备所产生的上述提及的受控频率波形是直接的数字合成。 

2 直接数字频率合成器 

一个典型的直接数字频率合成器包含一个正弦波(正弦查找表 LUT)样品 RAM。在限定相位跳跃的频率设置字的控制方式下来搜寻这些样本。一个典型 的频率设置字是 32 位宽,但 48 位合成器在较高的频率分辨率下也可使用。一个 相位累加器产生连续的正弦查找表的地址,并生成一个数字正弦波输出。DDS 的数字部分,即相位累加器和查表,被称为数控振荡器(NCO) 。最后阶段,这相对于前一个主要是模拟的,包括一个 D / A 转换器和一个过滤器。过滤器使 数字化的正弦波更平稳,产生连续输出信号。在凡需要方波输出的应用中,这由一个硬器在经过过滤器之后得到。这不等于使用例如累加器的输出,硬过滤和波形输出也不是最高位有限,因为会遇到显著的抖动。

 对于 n 位系统的输出信号的频率是按以下方式计算的;如果相步进n等于 1,将累 加器的计数加 1,需要2个时钟周期,以满足整个 LUT 生成一个周期的正弦波。 这是该系统能生成的最低的频率,也是它的频率分辨率。设置字 FSW 为二,计数 器的结果间隔数为二,需要个时钟周期来完成一个周期的正弦波输出。它可以很容易地被证明,对于任意整数 m(其中) ,所采取的时钟周期数旨在产生一个输出的正弦波,周期为,输出频率(fDDS)和频率分辨率(fres)由下列公式给出:

                                           

                                           

对于 n = 32,有一个 fclk = 33 MHz 的时钟频率,频率分辨率为 7.68 兆赫兹。如 果 n 增加至48个相同的时钟频率,分辨率为 120 nHz 是可能的。 

3 提出的频率测量技术 

我们目前的设计想法来自于具有极高的DDS 频率分辨率的设备并且由它的封闭循环的抗干扰形式执行。一个(已知)频率源,即 DDS,采用一个闭环 ,并且被迫逐步产生频率等于未知输入的输出。DDS 系统的经验法则 所接受的最大合成频率为时钟频率的 25%(远低于奈奎斯特) 。根据这一理论,我们的原型是使用一个 33 MHz 的时钟将有效数定为 8 兆赫。在砷化镓产品来看, 我们可以看到,最近获得的 DDS 设计可以在高达 400 兆赫的时钟频率范围运作。因此根据目前的方法,工作频率高达 100 MHz的频率计数器是可以设计的。该决议将取决于 FSW 的数量和时钟频率。 DDS 的时钟频率是非常重要的,因为随着它的减小, 该方法(定义为 fclk /)的结果会变的更加完美。时钟频率下降影响的是其最大输出频率 即计数器的最大计数随之降低。 本文中主要模块已被证明。 其中包括:频率比较器和 DDS。为了克服特定频率比较器的一些缺点它已纳入校正阶段。这一阶段也可用于测量提取,以显示正确的读数。 

3.1 电路的操作 

该电路工作在这样的方式即一个新测量 的DDS 的输出频率会在开始以逐次逼近的方法受到控制 。初始 DDS 的频率将是它最大值的一半。此外,该步骤将频率 近似等于 DDS 的最大频率的 1/ 4。根据比较器输出的频率,在每一个近似值中 频率被分成两个并且增加或减少到 DDS 的 FSW 中。在步长下降到一定程度时逼近过程停止。在此之后,向上/向下计数器替代逼近机制。

 在适当的修正和解码后,数码的 FSW 被显示在一个输出设备中,即一台液晶 显示器或任何其他合适的设备。或者,也可以进行数字记录,也可以由计算机阅 读。

 作为这一初步方法的结论,我们可以说,该方法是基于数字控制合成的,这个数字合成器能被迫产生和未知频率几乎相等的频率。 

3.2 频率比较 

频率比较似乎是设计的最关键阶段。该实现是基于一种改进的相位/频率比 较器,由飞利浦在 74HC4046 PLL 设备中生产。它主要包括两个二进制计数器和一个 RS 触发器。

 频率比较器的功能是基于较低频率,即较大的周期的原则,包括至少有 一个或多个频率较高(小周期)的完整周期。这意味着,在较低频率周期内包含两个或两个以上的有较高频 率上升边缘的波形。鉴于上述情况,电路操作如下:当第一个计数器(#1)在一个时期内遇到 DDS 的两个未知频率的上升边缘,它设置 RS 触发器的输出。 触发器的逻辑“1”在向上/向下计数器的 U / D 的控制输出中起 RS 作用,强制 DDS 升高输出频率。相反,当第二个计数器(#2)在一个周期内记录DDS输出的两个未知频率的上升边缘,它又恢复成RS 触发器的输出 。这个动作降低了 DDS 的频率。

 乍一看人们可能认为,合成频率可达到实测值,然后计数器停止运作。不幸 的是并非如此。一个充满活力的机制代替了它。该电路需要一些时间来实现正确的频率关系。我们将把这个时间称为“迟滞” 。迟滞取决于最初的 DDS 输出时序关系和未知频率。最初,在滞后期,有关更大的频率的指示是不明确的,即它可能是错误的。当更高的频率上升边缘波形发生在较低的时期时将会产生歧义。当我们考虑 DDS 的频率等于未知频率之一时,我们会发现,比较器的输出将切换,说明 DDS 的频率高于或低于未知频率。这实际上是一个可以接受和预期的状况,因为一个相同的指示(如电压比较器)可能不存在。在我们 的例子中,这不是一个问题,因为这个电路是嵌入在一个封闭的循环之中。该循环经过一段短暂的时间,迟滞等情况将得到扭转。滞后的时间是可变的。这种情况如何控制,也将在后面解释。虽然与模拟执行频率相比较将产生更大的噪音,我们仍坚持数字化的实现,原因有三:超大型电路或可编程逻辑器件(PLD)容易实现,不需要模拟组件,频率范围宽并且需要更短的响应时间。 

3.3 频率比较器和数字频率合成器之间的相互作用 

在频率比较器“实现” 的未知频率逐次逼近之后,合成的频率较高(低)于未知, 并在控制向上/向下计数器的输出端产生向下(上)的一个逻辑 0(1)的方向。 如前所述,这个计数器的输出被认为是从 FSW 到 DDS 的阶段。在最初的 DDS 频率降低时,合成频率将会逐步增加,达到未知频率之一。这不会通过频率比较器“实 现”,合成频率将会在一些时钟周期内继续增加,直到比较器检测出它的两个输入即频率未知的一方和 DDS 输出的正确关系。在相反(降低)的情况下,同样的 现象也将会被观察到。这是因为前面提到的滞后作用。

 当 DDS 输出(fDDS)已接近fin,由于滞后性,没有特定的频率合成。相反,它 摇摆于 F1 和 F2 之间,其中 F1 和 F2 是频率对称摆动的两个极端值。 DDS 的 输出可以被看作是一个三角波形调制载波频率。三角波形是 FSW 施加到 DDS 的模拟表示法。较低的跟踪显示一个比较典型的频率输出。在相同的图上跟踪,  以模拟的形式显示 FSW 的变化, 这是因为它企图接近正确的值。 利用辅助硬件电路这个波形已被俘获:为了研究操作,输出数字至模拟转换器(DAC)连接到 U / D 转换计数器(最高位) 。该 DAC 没有显示电路框图。换句话说,较低水平的跟踪是将 U / D 命令(输入)到计数器上,而跟踪的痕迹是一个假设的“调频”波形。很明显,使用“假设”是因为在电路(除辅助 DAC)中没有一个可用的波形。相反,其相等数值存在。三角波形的坡度大小对于常数输入频率是 恒定的,并且取决于 U/ D 转换计数器(水平轴)时钟和 DAC(垂直轴)的电压基准。 这里的坡度为 ±。

 3.4硬件原型的说明 

用于评估的两个原型在实验室已建成并进行测试。 第一种方法是用低频率仪器 (工 作达 15 千赫) 。这次实验实施的目的是研究该方法的操作原则。接下来,用一个更高频率的仪器原型,将在这里进行更详细的描述。为了使原型的数字部分(频率 比较,连续计数器,校正阶段)生效,两个产自 Altera(EPF80LC68 - 12) 的 PLD 器件将被使用。 这些设备和由高通 Q2240I - 3S1 所生产 DDS 相互联系。 DDS 具有 32 位输入和12 位输出的正弦查找表 (LUT) 。 送入到由模拟设备 AD9713B 发出的 D / A 转换器中。其模拟输出连接到 I / V 放 大器(电流电压转换器) 。 

由于 DAC 的工作,生成的正弦波具有较高的谐波。这些谐波在 DAC 之后将从过 滤器删除。这次调整阶段一部分实施在 PLD 一部分在微控制器。基于频率比较 器的上下命令,我们存储两个极端值,FSW1 和 FSW2,然后再进入微控制器( Atmel ATC52) 转换成数字表示并反馈到 LCD 显示器。 该微控制器还控制着整个运作的原型。 

仪器的行为和预期的一样,和常规的频率计数器工作台是一样的。在数字示波器的帮助下,测量采用较低速度跟踪检查。每个状态,波形的高或低,对应一个测量所需的时间。

 4 结论 

在该文件中频率测量的替代方法已经提出。据指出,在大多数情况下,对于相同频率的解决方案,这种方法比传统方法更快。另一方面,由于 DDS 固有的高频率的特点,该方法的精度非常高。这种可作为振荡器的合成器,在未知的输 入频率范围被驱使“振荡”。与常规方法相比,两个原型已建成并在实验室测试。 

这种方法的第二个主要优点是,如果重复频率测量,工具一旦锁定,频率测量不重新从头开始,而是自动驱使到更低或更高的值。换句话说,循环有能力按照输入信号频率的变化而改变。在传统的计算技术里,计算过程为每个新的测量量而重 复(重新启动) 。

 另一个重要优势是由于其闭环的性质,该系统具有很强的抗干扰能力。一个详细的对于噪音的研究已经在本文中指出。 这主要是因为本文的目的旨在提出一个频率测量的替代原理。此外,该系统的最终输出采取了进一步的(测量校正)有助于提高抗噪声能力的后处理。

文档

基于DDS的精密频率测量的一种替代方法

AnalternativemethodofprecisefrequencybytheaidofaDDSContentsAmethodoffrequencymeasurementbasedonaclosedloopcomposedmainlyofaFrequencyComparator(FC)andaDirectDigitalSynthesizer(DDS)ispresentedinthispaper.TheDDSservesasreferencesinewavesignalgeneratora
推荐度:
  • 热门焦点

最新推荐

猜你喜欢

热门推荐

专题
Top