最新文章专题视频专题问答1问答10问答100问答1000问答2000关键字专题1关键字专题50关键字专题500关键字专题1500TAG最新视频文章推荐1 推荐3 推荐5 推荐7 推荐9 推荐11 推荐13 推荐15 推荐17 推荐19 推荐21 推荐23 推荐25 推荐27 推荐29 推荐31 推荐33 推荐35 推荐37视频文章20视频文章30视频文章40视频文章50视频文章60 视频文章70视频文章80视频文章90视频文章100视频文章120视频文章140 视频2关键字专题关键字专题tag2tag3文章专题文章专题2文章索引1文章索引2文章索引3文章索引4文章索引5123456789101112131415文章专题3
当前位置: 首页 - 正文

MLSS测定国标

来源:动视网 责编:小OO 时间:2025-10-05 05:19:39
文档

MLSS测定国标

摘自—水和废水监测分析方法(第四版)TotalSuspendedSolidsDriedat103–105°C——摘自美国国家标准《Standard_Methods_for_the_Examination_of_Water_and_Wastewater20thEdition》1.GeneralDiscussiona.Principle:Awell-mixedsampleisfilteredthroughaweighedstandardglass-fiberfilterandtheresiduer
推荐度:
导读摘自—水和废水监测分析方法(第四版)TotalSuspendedSolidsDriedat103–105°C——摘自美国国家标准《Standard_Methods_for_the_Examination_of_Water_and_Wastewater20thEdition》1.GeneralDiscussiona.Principle:Awell-mixedsampleisfilteredthroughaweighedstandardglass-fiberfilterandtheresiduer
摘自—水和废水监测分析方法 (第四版)

Total Suspended Solids Dried at 103–105°C

——摘自美国国家标准《Standard_Methods_for_the_Examination_of_Water_and_Wastewater  20th Edition》

1.  General Discussion

a.  Principle: A well-mixed sample is filtered through a weighed standard glass-fiber filter

and the residue retained on the filter is dried to a constant weight at 103 to 105°C. The increase

in weight of the filter represents the total suspended solids. If the suspended material clogs the

filter and prolongs filtration, it may be necessary to increase the diameter of the filter or decrease

the sample volume. To obtain an estimate of total suspended solids, calculate the difference

between total dissolved solids and total solids.

b.  Interferences: See Section 2540A.2 and  Section 2540B.1. Exclude large floating particles

or submerged agglomerates of nonhomogeneous materials from the sample if it is determined

that their inclusion is not representative. Because excessive residue on the filter may form a water-entrapping crust, limit the sample size to that yielding no more than 200 mg residue. For

samples high in dissolved solids thoroughly wash the filter to ensure removal of dissolved

material. Prolonged filtration times resulting from filter clogging may produce high results

owing to increased colloidal materials captured on the clogged filter.

2.  Apparatus

A.Muffle furnace for operation at 550°C.

B.Desiccator, provided with a desiccant containing a color indicator of moisture

C.concentration or an instrumental indicator.

D.Drying oven, for operation at 103 to 105°C.

E.Analytical balance, capable of weighing to 0.1 mg.

F.Magnetic stirrer with TFE stirring bar.

G.Wide-bore pipets.#(45)†

H.Graduated cylinder.

I.Low-form beaker.#(46)‡ 

J.Glass-fiber filter disks#(47)* without organic binder.

K.Filtration apparatus: One of the following, suitable for the filter disk selected:

1)  Membrane filter funnel. 

2)  Gooch crucible, 25-mL to 40-mL capacity, with Gooch crucible adapter. 

3)  Filtration apparatus with reservoir and coarse (40- to 60-µm) fritted disk as filter

support.#(48)† 

L.Suction flask, of sufficient capacity for sample size selected.

M.  Aluminum weighing dishes. 

3.  Procedure

a.  Preparation of glass-fiber filter disk: If pre-prepared glass fiber filter disks are used,

eliminate this step. Insert disk with wrinkled side up in filtration apparatus. Apply vacuum and

wash disk with three successive 20-mL portions of reagent-grade water. Continue suction to

remove all traces of water, turn vacuum off, and discard washings. Remove filter from filtration

apparatus and transfer to an inert aluminum weighing dish. If a Gooch crucible is used, remove

crucible and filter combination. Dry in an oven at 103 to 105°C for 1 h. If volatile solids are to

be measured, ignite at 550°C for 15 min in a muffle furnace. Cool in desiccator to balance

temperature and weigh. Repeat cycle of drying or igniting, cooling, desiccating, and weighing

until a constant weight is obtained or until weight change is less than 4% of the previous

weighing or 0.5 mg, whichever is less. Store in desiccator until needed.

b.  Selection of filter and sample sizes: Choose sample volume to yield between 2.5 and 200

mg dried residue. If volume filtered fails to meet minimum yield, increase sample volume up to 1

L. If complete filtration takes more than 10 min, increase filter diameter or decrease sample

volume.

c.  Sample analysis: Assemble filtering apparatus and filter and begin suction. Wet filter with

a small volume of reagent-grade water to seat it. Stir sample with a magnetic stirrer at a speed to

shear larger particles, if practical, to obtain a more uniform (preferably homogeneous) particle

size. Centrifugal force may separate particles by size and density, resulting in poor precision

when point of sample withdrawal is varied. While stirring, pipet a measured volume onto the

seated glass-fiber filter. For homogeneous samples, pipet from the approximate midpoint of

container but not in vortex. Choose a point both middepth and midway between wall and vortex.

Wash filter with three successive 10-mL volumes of reagent-grade water, allowing complete

drainage between washings, and continue suction for about 3 min after filtration is complete.

Samples with high dissolved solids may require additional washings. Carefully remove filter

from filtration apparatus and transfer to an aluminum weighing dish as a support. Alternatively,

remove the crucible and filter combination from the crucible adapter if a Gooch crucible is used.

Dry for at least 1 h at 103 to 105°C in an oven, cool in a desiccator to balance temperature, and

weigh. Repeat the cycle of drying, cooling, desiccating, and weighing until a constant weight is

obtained or until the weight change is less than 4% of the previous weight or 0.5 mg, whichever

is less. Analyze at least 10% of all samples in duplicate. Duplicate determinations should agree within 5% of their average weight. If volatile solids are to be determined, treat the residue

according to 2540E.

4.  Calculation

 

where: 

                  A =   weight of filter + dried residue, mg, and 

                  B =   weight of filter, mg. 

5.  Precision

 The standard deviation was 5.2 mg/L (coefficient of variation 33%) at 15 mg/L, 24 mg/L

(10%) at 242 mg/L, and 13 mg/L (0.76%) at 1707 mg/L in studies by two analysts of four sets of

10 determinations each. 

Single-laboratory duplicate analyses of 50 samples of water and wastewater were made with

a standard deviation of differences of 2.8 mg/L.